Comparison of absorbed doses resulting from various intraoral periapical radiography
Creators
- 1. Dept. of Oral and Maxillofacial Radiology, Graduate School, Seoul National University, Seoul (Korea, Republic of)
Description
This study was designed to measure the absorbed dose to organs of special interest from full mouth with intraoral film (14 films) and to compare the five periapical techniques. Thermoluminescent crystals (TLD-100 chip) were located in brain, orbit, bone marrow of mandibular ramus, bone marrow of mandibular body, bone marrow of 4th cervical spine, parotid gland, submandibular gland and thyroid gland. X-ray machine was operated at 70 kVp and round collimating film holding device (XCP) and rectangular collimating film holding device (Precision Instrument) were used. The distance from the X-ray focus to the open end of the collimator was 8 inch, 12 inch and 16 inch. The following results obtained; 1. The absorbed dose was the highest in bone marrow of mandibular body (5.656 mGy) and the lowest in brain (0.050 mGy). 2. Generally, the lowest absorbed dose was measured from 16 inch cylinder, rectangular collimating film holding device with paralleling technique. But, in bone marrow of mandibular body and the floor of mouth, the highest absorbed dose was measured from 12 inch cylinder, rectangular collimating film holding device with paralleling techniques. 3. Comparing of five intraoral radiographic techniques, it was appeared statistically significant reduction of the absorbed doses measured with rectangualr collimating film holding device compared to XCP film holding device (p<0.05). 4. No statistically significant reduction in the absorbed dose was found as cylinder length was change (p>0.05).
Additional details
Publishing Information
- Journal Title
- Korean Journal of Oral and Maxillofacial Radiology
- Journal Volume
- 25
- Journal Issue
- 2
- Series
- 27 refs, 1 fig, 11 tabs
- Journal Page Range
- p. 297-308
- ISSN
- 1229-8212
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 43010833
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BONE MARROW; DESIGN; IMAGES; JAW; RADIATION DOSES
- Descriptors DEC
- ANIMAL TISSUES; BODY; DOSES; HEMATOPOIETIC SYSTEM; ORGANS; SKELETON; SKULL