Published June 2016 | Version v1
Journal article

The Dortmund Low Background Facility — Low-background gamma ray spectrometry with an artificial overburden

  • 1. Technische Universität Dortmund, Lehrstuhl für Experimentelle Physik IV, Otto-Hahn-Str. 4, 44221 Dortmund (Germany)
  • 2. Technische Universität Dresden, Institut für Kern-und Teilchenphysik, Zellescher Weg 19, 01069 Dresden (Germany)

Description

The Dortmund Low Background Facility is an instrument for low-level gamma ray spectrometry with an artificial overburden of ten meters of water equivalent, an inner shielding, featuring a neutron absorber, and an active muon veto. An integral background count rate between 40 keV and 2700 keV of (2.528±0.004)counts/(kg min) enables low-background gamma ray spectrometry with sensitivities in the range of some 10 mBq/kg within a week of measurement time.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2016.03.025

Additional details

Identifiers

DOI
10.1016/j.apradiso.2016.03.025;
PII
S0969-8043(16)30118-X;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
112
Journal Page Range
p. 165-176
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.