Comparison of SAR calculation algorithms for the finite-difference time-domain method

  • 1. Department of Radio Science and Engineering, Aalto University, Otakaari 5 A, 02150 Espoo (Finland)

Description

Finite-difference time-domain (FDTD) simulations of specific-absorption rate (SAR) have several uncertainty factors. For example, significantly varying SAR values may result from the use of different algorithms for determining the SAR from the FDTD electric field. The objective of this paper is to rigorously study the divergence of SAR values due to different SAR calculation algorithms and to examine if some SAR calculation algorithm should be preferred over others. For this purpose, numerical FDTD results are compared to analytical solutions in a one-dimensional layered model and a three-dimensional spherical object. Additionally, the implications of SAR calculation algorithms for dosimetry of anatomically realistic whole-body models are studied. The results show that the trapezium algorithm-based on the trapezium integration rule-is always conservative compared to the analytic solution, making it a good choice for worst-case exposure assessment. In contrast, the mid-ordinate algorithm-named after the mid-ordinate integration rule-usually underestimates the analytic SAR. The linear algorithm-which is approximately a weighted average of the two-seems to be the most accurate choice overall, typically giving the best fit with the shape of the analytic SAR distribution. For anatomically realistic models, the whole-body SAR difference between different algorithms is relatively independent of the used body model, incident direction and polarization of the plane wave. The main factors affecting the difference are cell size and frequency. The choice of the SAR calculation algorithm is an important simulation parameter in high-frequency FDTD SAR calculations, and it should be explained to allow intercomparison of the results between different studies. (note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/55/15/N03

Additional details

Identifiers

DOI
10.1088/0031-9155/55/15/N03;
PII
S0031-9155(10)50825-X;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
55
Journal Issue
15
Journal Page Range
p. N421-N431
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42030633
Subject category
S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
Descriptors DEI
ALGORITHMS; DOSIMETRY; ENERGY ABSORPTION; FINITE DIFFERENCE METHOD; SIMULATION
Descriptors DEC
ABSORPTION; CALCULATION METHODS; ITERATIVE METHODS; MATHEMATICAL LOGIC; MATHEMATICAL SOLUTIONS; NUMERICAL SOLUTION; SORPTION