Semi-empirical gamma-ray response function of BGO, NaI(Tl) and LaBr 3 (Ce) scintillation detectors
Creators
- 1. Institute for Nuclear Research and Nuclear Energy (INRNE), BAS, Sofia (Bulgaria)
- 2. Joint Institute for Nuclear Research (JINR), Dubna (Russian Federation)
- 3. Faculty of Physics, Lomonosov Moscow State University (MSU), Moscow (Russian Federation)
- 4. Skobeltsyn Institute of Nuclear Physics (SINP), MSU, Moscow (Russian Federation)
- 5. The Institute of Nuclear Physics, Ministry of Energy of the Republic of Kazakhstan, 050032 Almaty (Kazakhstan)
- 6. L.N. Gumilyov Eurasian National University, 010008 Nur-Sultan (Kazakhstan)
Description
The response function of the BGO, NaI (Tl) and LaBr(Ce) scintillation detectors to monoenergetic gamma quanta was built on the basis of Monte Carlo simulations using the GEANT4 toolkit and calibration measurements with gamma radiation sources of different energies. The response function consists of seven components: the first six constitute the detector's response to a direct hit of monoenergetic gamma radiation and depend on the properties of the detector (size, material, energy resolution, etc.), and the seventh component represents the effect of surrounding materials on the measured gamma spectrum. For each component of the function, the analytical form of the energy dependence is determined and its parameters are found when registering gamma quanta with energies in the range from 0.3 MeV to 10 MeV.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2021.165741Additional details
Identifiers
- DOI
- 10.1016/j.nima.2021.165741;
- PII
- S0168900221007269;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 1014
- Journal Page Range
- vp.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54084077
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S73: NUCLEAR PHYSICS AND RADIATION PHYSICS;
- Descriptors DEI
- CALIBRATION; COMPUTERIZED SIMULATION; ENERGY DEPENDENCE; ENERGY RESOLUTION; GAMMA RADIATION; GAMMA SPECTRA; MEV RANGE; MONTE CARLO METHOD; RADIATION SOURCES; RESPONSE FUNCTIONS; SCINTILLATION COUNTERS; SODIUM IODIDES
- Descriptors DEC
- ALKALI METAL COMPOUNDS; CALCULATION METHODS; ELECTROMAGNETIC RADIATION; ENERGY RANGE; FUNCTIONS; HALIDES; HALOGEN COMPOUNDS; INORGANIC PHOSPHORS; IODIDES; IODINE COMPOUNDS; IONIZING RADIATIONS; MEASURING INSTRUMENTS; PHOSPHORS; RADIATION DETECTORS; RADIATIONS; RESOLUTION; SIMULATION; SODIUM COMPOUNDS; SODIUM HALIDES; SPECTRA
Optional Information
- Copyright
- Copyright (c) 2021 Elsevier B.V. All rights reserved.
- Collaborations
- TANGRA collaboration