Description

Based on experimentally obtained angular response functions a simple analytic function has been found which can be accurately fitted to such data. Using this function formulas are derived for the variation of the detector response as the angular distribution of the electrons are changing with depth in a phantom both inelectron and photon beams. The formulas show that for a flat liquid ionization chamber, a response variation of 2 to 4% is normally obtained in the build-up region of electron and photon beams whereas the response is practically constant beyond dose maximum. (orig.)

Additional details

Publishing Information

Journal Title
Acta Radiol., Oncol.
Journal Volume
24
Journal Issue
3
Series
Acta Radiol., Oncol.
Journal Page Range
301-304
ISSN
0349-652X
CODEN
AROBD