Contrast effects of a gadolinium filter
Creators
- 1. Department of Radiology, University of British Columbia, Vancouver, Canada
Description
Several authors have suggested using heavy metal filters with K edges in the diagnostic energy range to reduce the width of the x-ray spectrum and hence reduce patient radiation exposure. This spectral narrowing also increases subject contrast and permits an increase in tube potential. Results of contrast measurements are presented for a 250 mu gadolinium filter. It was found that aluminum filter contrast could be matched by using 8 to 10 kVp higher potential with the gadolinium filter. Similar results were found for calcium tungstate and rare-earth screens. Measurements were also done to determine skin exposure and mAs ratios for both constant contrast and constant kVp technique conversion methods. A simple theory with one adjustable parameter gives a reasonable fit to the experimental results
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 8
- Journal Issue
- 2
- Series
- Med. Phys.
- Journal Page Range
- 203-9
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14719702
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPARATIVE EVALUATIONS; DIAGNOSIS; FILTERS; GADOLINIUM; PATIENTS; RADIATION DOSES; RADIATION HAZARDS; SKIN; SPECIFICATIONS; X RADIATION
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; ELEMENTS; HAZARDS; HEALTH HAZARDS; IONIZING RADIATIONS; MEDICINE; METALS; ORGANS; RADIATIONS; RARE EARTHS