Published February 1986 | Version v1
Journal article

TL and TSEE dosemeter of Cu+ implanted Li2B4O7 glass ceramics

  • 1. Government Industrial Research Inst., Osaka, Ikeda (Japan)

Description

Simultaneous TL and TSEE measurements were carried out with Li2B4O7 glass ceramics in which Cu+ had been implanted and glass ceramics which were made of (Li2B4O7 : Cu + SiO2 + Pt). For all cases it was found that the TSEE responses were stable and reliable. However, the TL response of the Cu+-implanted samples showed a very unstable behavior, probably because the implanted Cu+ could not find stable sites in crystals under the heat treatment condition adopted in this experiment. It was also found that (Li2B4O7 : Cu + SiO2 + Pt) glass ceramics had good TL and TSEE responses as solid-state dosemeters. (author)

Additional details

Publishing Information

Journal Title
Jpn. J. Appl. Phys., Suppl.
Journal Volume
24
Journal Issue
4
Series
Jpn. J. Appl. Phys., Suppl.
Journal Page Range
242-245
ISSN