This document presents a tutorial description of several concepts and definitions of limits of quantitation, such as the so-called 10 s, the RSDnet, the method limit, the lower limit of the calibration graph and the uncertainty approaches. Use, advantages, limitations and complexity of the various approaches are illustrated with some examples taken from atomic spectrometry, using ICP-AES, ICP-MS and LIBS. Information that can be deduced from the calibration graph is emphasized

Availability note (English)

Available from http://dx.doi.org/10.1016/j.sab.2007.11.029

Additional details

Identifiers

DOI
10.1016/j.sab.2007.11.029;
PII
S0584-8547(07)00443-0;

Publishing Information

Journal Title
Spectrochimica Acta. Part B, Atomic Spectroscopy
Journal Volume
63
Journal Issue
2
Journal Page Range
p. 166-182
ISSN
0584-8547
CODEN
SAASBH

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
39085487
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
CALIBRATION; EMISSION SPECTROSCOPY; ICP MASS SPECTROSCOPY; MANUALS
Descriptors DEC
DOCUMENT TYPES; MASS SPECTROSCOPY; SPECTROSCOPY