A quality and efficiency analysis of the IMFASTTM segmentation algorithm in head and neck 'step and shoot' IMRT treatments
- 1. Department of Radiation Oncology, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514 (United States)
Description
The performance of segmentation algorithms used in IMFAST for 'step and shoot' IMRT treatment delivery is evaluated for three head and neck clinical treatments of different optimization objectives. The segmentation uses the intensity maps generated by the in-house TPS PLANUNC using the index-dose minimization algorithm. The dose optimization objectives include PTV dose uniformity and dose volume histogram-specified critical structure sparing. The optimized continuous intensity maps were truncated into five and ten intensity levels and exported to IMFAST for MLC segments optimization. The MLC segments were imported back to PLUNC for dose optimization quality calculation. The five basic segmentation algorithms included in IMFAST were evaluated alone and in combination with either tongue and groove/match line correction or fluence correction or both. Two criteria were used in the evaluation: treatment efficiency represented by the total number of MLC segments and optimization quality represented by a clinically relevant optimization quality factor. We found that the treatment efficiency depends first on the number of intensity levels used in the intensity map and second the segmentation technique used. The standard optimal segmentation with fluence correction is a consistent good performer for all treatment plans studied. All segmentation techniques evaluated produced treatments with similar dose optimization quality values, especially when ten-level intensity maps are used
Additional details
Identifiers
- DOI
- 10.1118/1.1428755;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 29
- Journal Issue
- 3
- Journal Page Range
- p. 275-283
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35004175
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BEAM POSITION; DOSIMETRY; IMAGE PROCESSING; OPTIMIZATION; RADIOTHERAPY
- Descriptors DEC
- MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY; THERAPY
Optional Information
- Notes
- (c) 2002 American Association of Physicists in Medicine.