Published August 21, 1999 | Version v1
Journal article

First attempts to combine capillary tubes with photocathodes

Description

We describe our efforts to combine glass capillary plates, operating as a gas amplification structure at ∼1 atm, with photocathodes sensitive to visible light. Such capillary tubes are a by-product of the manufacture of Microchannel Plates and are commercially available. Preliminary tests indicate that gas gains >103 could be achieved without photon feedback. With two capillary plates in tandem (double-step multiplication) overall gains up to 105 were possible at counting rate <100 Hz/mm2. This approach may open new possibilities for detection of visible photons by gaseous detectors. Potential advantages are: high gains, large sensitive area, high granularity, and insensitivity to magnetic fields

Additional details

Identifiers

PII
S0168900299003460;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
433
Journal Issue
1-2
Journal Page Range
p. 492-501
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
Bulgaria
INIS RN
34051253
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
AMPLIFICATION; GLASS; PHOTOCATHODES; PHOTOTUBES; SENSITIVITY; VISIBLE RADIATION
Descriptors DEC
CATHODES; ELECTRODES; ELECTROMAGNETIC RADIATION; RADIATIONS

Optional Information

Copyright
Copyright (c) 1999 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.