Published February 1, 1998 | Version v1
Journal article

Novel electrode geometry to improve performance of CdZnTe detectors

  • 1. Danish Space Res. Inst., Copenhagen (Denmark)

Description

A novel electrode geometry to improve the sensitivity to single-polarity charge carriers in ionization detectors is presented. The electrode geometry is based on the same principle as Frisch grids in gas detectors. It reduces the sensitivity to one type of charge carrier, making the detector almost unipolar sensing. The effectiveness of the geometry was demonstrated with a CdZnTe crystal. Above X-ray energies of 40 keV, large improvements were achieved in the spectral sensitivity, and the spectral resolution could be increased further by applying a correction for charge trapping. The peak of 137Cs at 661 keV was not visible with a planar electrode, while its FWHM was 16.7% with the novel electrode geometry. The FWHM was 6.1% after correction for hole trapping. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
403
Journal Issue
2-3
Journal Page Range
p. 390-398.
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Notes
8 refs.