Published July 2013 | Version v1
Journal article

Commissioning and quality assurance of Calypso four-dimensional target localization system in linear accelerator facility

  • 1. Department of Radiation Oncology, American Oncology Institute, Hyderabad (India)

Description

Four dimensional (4D) target localization system (Calypso System) was installed at our hospital, which is equipped with Beacon Transponders, Console, Electromagnetic Array, Optical System, Tracking Station, Treatment table overlay, and Calypso kVue Couch top. The objective of this presentation is to describe the results of commissioning measurements carried out on the Calypso System to verify the manufacturer specifications and also to evolve a quality assurance (QA) procedure which can be used to test its performance routinely. The QA program consists of a series of tests (QA for checking the calibration or system accuracy, Camera Calibration with L-frame fixture, Camera Calibration with T-frame fixture, System calibration Fixture targets test, Localization, and Tracking). These tests were found to be useful to assess the performance of the Calypso System. (author)

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
38
Journal Issue
3
Journal Page Range
p. 143-147
CODEN
JMPHFE

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
44120916
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
ACCURACY; LINEAR ACCELERATORS; QUALITY ASSURANCE; RADIOTHERAPY; REAL TIME SYSTEMS
Descriptors DEC
ACCELERATORS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Notes
11 refs., 6 figs., 1 tab.