Published 1985 | Version v1
Report Open

Thirty-channel information and measuring system for determining the KVTI-20 accelerator synchrotron radiation power and angular distribution

Description

The thirty channel system for measuring the synchrotron radiation power and angular distribution in the KVTI-20 accelerator is described. The system allows one to detect the radiation in 1-4 μm spectral region for ten time momenta during 1. ms, when the electron ring is compressed from 5.2 to 3.8 cm radii. The detector consists of uncoold phoresistors made of PbSe with dimensions 0.2x4 mm with linear arrangement of elements 120 mm long. The system is intended for the investigation of the electronion ring dynamics during compression in the KVTI-20 adgezator

Availability note (English)

MF available from INIS under the Report Number.

Files

17036958.pdf

Files (118.1 kB)

Name Size Download all
md5:a510d4bfc8c3f7b13bd1a549ef7b545f
118.1 kB Preview Download

System files (21.2 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
30-канальная информационно-измерительная система для определения интенсивности и углового распределения синхротронного излучения KUTI-20

Publishing Information

Imprint Pagination
4 p.
Report number
JINR--9-85-278

Optional Information

Notes
6 refs.; 2 figs.