Published October 2011
| Version v1
Journal article
Optimization of MOSFET calibration for in vivo dosimetry in radiosurgery: reduction of measurement uncertainties in pre-clinical conditions
Creators
- 1. UMR 825 'imagerie cerebrale et handicaps neurologiques', Inserm, Toulouse (France)
- 2. Unite de radiophysique et de radioprotection, CHU de Toulouse, Toulouse (France)
- 3. Best medical Canada, Ottawa (Canada)
- 4. Centre de radiochirurgie stereotaxique, CHU Rangueil, Toulouse (France)
- 5. Groupe Oncorad Garonne, Montauban (France)
Description
The objective of this study is to assess the conventional formulas of equivalent square for fields with irregular geometry, by transposing the optimized calibration method which has been previously developed, to micro-MOSFET. The study has been performed on a 6 MV Novalis apparatus equipped with micro-multi-blades collimators (BrainLab). The average dose bias reaches 2.66 per cent for all field sizes. Therefore, it appears that the joint use of the square inverse of distances and of conventional formulas of equivalent square results in an acceptable in vivo dosimetry precision. Short communication
Availability note (English)
Available from doi: http://dx.doi.org/10.1016/j.canrad.2011.07.160Additional details
Additional titles
- Original title (French)
- Optimisation de la calibration de MOSFET pour la dosimetrie in vivo en radiochirurgie: reduction des incertitudes de mesure en conditions precliniques
Identifiers
Publishing Information
- Journal Title
- Cancer Radiotherapie
- Journal Volume
- 15
- Journal Issue
- 6-7
- Journal Page Range
- p. 617-618
- ISSN
- 1278-3218
- CODEN
- CARAFC
Conference
- Title
- 22. national congress of the French society of oncological radiotherapy (SFRO)
- Original Conference Title
- 22. congres national de la Societe Francaise de Radiotherapie Oncologique (SFRO)
- Dates
- 5-7 Oct 2011
- Place
- Paris (France)
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 43041620
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S97: MATHEMATICAL METHODS AND COMPUTING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; CALCULATION METHODS; CALIBRATION; COLLIMATORS; DATA COVARIANCES; DOSIMETRY; IN VIVO; MOSFET; NEOPLASMS; OPTIMIZATION; RADIATION DOSES; RADIATION MONITORING; RADIOTHERAPY
- Descriptors DEC
- DISEASES; DOSES; FIELD EFFECT TRANSISTORS; MEDICINE; MONITORING; MOS TRANSISTORS; NUCLEAR MEDICINE; RADIOLOGY; SEMICONDUCTOR DEVICES; THERAPY; TRANSISTORS