Published September 2003 | Version v1
Journal article

MLC dosimetric characteristics for small field and IMRT applications

  • 1. University of Texas Medical Branch, Department of Radiation Oncology, Galveston, Texas 77555 (United States)

Description

The objective of this work was to measure the performance characteristics of a double-focus multileaf collimator (MLC) for intensity modulated radiation therapy (IMRT), specifically the variation in penumbra and leakage for narrow fields as a function of field position over a 20x27 cm space available for segmented MLC IMRT. Measurements were made with 6 MV x rays through a MLC containing 29 leaf pairs (27 pairs of 1 cm width), and EDR2 film at 10 cm depth in solid water at 100 cm SAD. Films were digitized with 0.17 mm resolution and converted to dose. Interleaf and intraleaf transmission were measured along 11 vertical profile locations. Leaf-end transmission was measured along horizontal profiles for each of 9 different leaf abutments, traveling over a 20 cm range. In-plane penumbra measurements were made through a single leaf retracted, for 7 different leaves. Cross-plane penumbra (leaf-end) measurements were made for all 27 leaf pairs, where the 1 cm field width was placed in 11 different off-axis positions (20 cm range). Interleaf leakage (range 1.0%-1.5%), intraleaf transmission (range 0.6%-0.8%), and leaf-end transmission (range 0.8%-2.7%) were consistent for all leaf pairs at a given abutment position. The penumbra for these 1-cm-wide fields was measured to be 0.36 cm±0.03 cm for 99% of the measurements. In conclusion, the penumbra and leakage of the double-focus MLC were remarkably consistent for the range of leaf positions studied, producing dosimetric characteristics that are well suited for IMRT segments where opposing leaf pairs are often separated by 10 mm or less

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
30
Journal Issue
9
Journal Page Range
p. 2545-2552
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35094420
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
COLLIMATORS; DOSIMETRY; FILMS; MODULATION; OPTICS; PERFORMANCE; RADIATION DOSES; RADIOTHERAPY; SPATIAL RESOLUTION
Descriptors DEC
DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; RESOLUTION; THERAPY

Optional Information

Notes
(c) 2003 American Association of Physicists in Medicine.