Published January 1979 | Version v1
Journal article

Radiation dose in abdominal examinations with whole body computer tomography

  • 1. Koeln Univ. (Germany, F.R.). Radiologisches Inst. und Poliklinik
  • 2. Zentralstelle fuer Sicherheitstechnik, Duesseldorf (Germany, F.R.)

Description

The somatic and genetic radiation exposure in abdominal CT examinations with Delta 50 and Delta 50 fast was measured via thermoluminescent dosimeters and compared with the exposure in conventional methods of X-ray diagnosis. With Delta 50 fast, the dose surface product for one examination of the pancreas is about 2 900 Rcm2, and for an examination of the kidneys about 3 400 Rcm2. It is thus comparable with a gastrointestinal examination or a urogram. The gonadal doses determined for CT are lower than those received from standard X-ray examinations. (orig.)

Abstract (German)

Es wurden die somatische und genetische Strahlenexposition bei abdominellen computertomographischen Untersuchungen am Delta 50 und Delta 50 fast mittels Thermolumineszenz-Dosimetern gemessen und mit der Strahlenbelastung herkoemmlicher roentgendiagnostischer Verfahren verglichen. Am Delta 50 fast betraegt das Flaechendosisprodukt fuer eine Untersuchung des Pankreas etwa 2900 Rcm2 und fuer eine Nierenuntersuchung etwa 3400 Rcm2 und ist mit einer Magen-Darm-Passage bzw. einem Urogramm vergleichbar. Die festgestellten Gonadendosen sind bei der Computertomographie geringer als bei den Standard-Roentgenuntersuchungen. (orig.)

Additional details

Additional titles

Original title (German)
Die Strahlenexposition bei abdominellen Untersuchungen mit der Ganzkoerper-Computertomographie

Publishing Information

Journal Title
Roentgen Bl.
Journal Volume
32
Journal Issue
1
Series
Roentgen Bl.
Journal Page Range
15-19