Published May 12, 2004 | Version v1
Journal article

Combined Imaging System for X-ray Fluorescence and Transmission X-ray Microtomography

  • 1. Institute of Applied Physics, University of Tsukuba, 1-1-1 Tennoudai, Tsukuba, Ibaraki 305-8573 (Japan)

Description

A full-field imaging x-ray fluorescence computed microtomography system was designed for quantitative elemental analysis. An x-ray fluorescence imaging microscope combined with transmission x-ray microtomographic system was constructed at BL3C2 of the Photon Factory, KEK, Japan. The quasi-monochromatic beam produced by a double-multilayer monochromator was used as an incident beam. X-ray fluorescence images were obtained and were reconstructed by simultaneous algebraic reconstruction technique (SART) with taking into account of the self-absorption effects of the incident beam and the x-ray fluorescence of the sample. Distribution of Fe in a standard sample (1.60 mol/l FeCl2 solution in a capillary tube) was observed to evaluate the system. The density of Fe reconstructed without correction of the self-absorption effect was 9.26 % of the actual density. Using this correction method, 73.0 % of the actual density could be obtained

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
705
Journal Issue
1
Journal Page Range
p. 1352-1355
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
8. international conference on synchrotron radiation instrumentation
Dates
25-29 Aug 2003
Place
San Francisco, CA (United States)

Optional Information

Notes
(c) 2004 American Institute of Physics