Published November 2019 | Version v1
Journal article

Gamma peak search and peak fitting algorithm for a low-resolution detector with applications in gamma spectroscopy

  • 1. Radiation Protection Bureau, Health Canada, Ottawa, ON (Canada)

Description

A Savitzky-Golay filtering for smoothing and peak search written in Python is presented in this paper alongside its applications in the list-mode digital data acquisition dual gamma-gamma coincidence bismuth germanate (BGO) detector. The study has demonstrated that the software provides a reliable and effective way to quantify trace amounts of 22Na and 7Be in aerosol samples collected at Resolute Bay, Canada with a critical limit of 3 mBq and 5 Bq respectively for a 20 h counting interval, which are believed to be the inherent limitations of the dual-BGO system. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
322
Journal Issue
2
Journal Page Range
p. 255-261
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
11 refs.