Published January 2001 | Version v1
Report Open

Source preparations for alpha and beta measurements

  • 1. Risoe National Lab., Roskilde (Denmark)

Description

Regarding alpha particle emitters subject for environmental studies, electrodeposition or co-precipitation as fluorides are the most common methods. For electro deposition stainless steel is generally used as cathode material but also other metals such as Ni, Ag, and Cu showed promising results. The use of other anode material than platinum, such as graphite should be investigated. For other purposes such as optimal resolution other more sophisticated methods are used but often resulting in poorer recovery. For beta particle emitters the type of detection system will decide the source preparation. Similar methods as for alpha particle emitters, electrodeposition or precipitation techniques can be used. Due to the continuous energy distribution of the beta pulse height distribution a high resolution is not required. Thicker sources from the precipitates or a stable isotopic carrier can be accepted but correction for absorption in the source must be done. (au)

Availability note (English)

Available from INIS in electronic form

Files

33036831.pdf

Files (207.7 kB)

Name Size Download all
md5:de0fb57000e4e93f1a6aff4051d3e13c
207.7 kB Preview Download

Additional details

Publishing Information

ISBN
87-7893-093-6
Imprint Pagination
17 p.
Report number
NKS--40

Optional Information

Contract/Grant/Project number
Contract NKS/BOK-1.1
Notes
2 tabls., 4 ills., 19 refs.