Communication and sampling rate limitations in IMRT delivery with a dynamic multileaf collimator system

  • 1. Department of Radiation Oncology, University of California at San Francisco, San Francisco, California 94143 (United States)

Description

The delivery of an intensity modulated radiation field with a dynamic multileaf collimator (MLC) requires precise correlation between MLC positions and cumulative monitor units (MUs). The purpose of this study is to investigate the precision of this correlation as a function of delivered MUs and dose rate. A semi-Gaussian shaped intensity profile and a simple geometric intensity pattern consisting of four square segments were designed to deliver a total of 1, 4, 16, 64, and 100 MUs at three different dose rates of 100, 400, and 600 MU/min. The semi-Gaussian intensity pattern was delivered using both sliding window and step and shoot techniques. The dose profiles of this intensity pattern were measured with films. The four square intensity pattern was delivered using step and shoot and conventional delivery techniques for comparison. Because of geometrical symmetry, the dose to each segment in this intensity pattern is expected to be the same when the same MU is assigned to each segment. An ionization chamber was used to measure the dose in the center of each of the four square segments. For the semi-Gaussian shaped profile, significant artifacts were observed when the profile was delivered with small MUs and/or at a high dose rate. For the four square intensity pattern, the dose measured in each segment presented a large variation when delivered with small MUs and a high dose rate. The variation increases as the MU/segment decreases and as the dose rate increases. These MU and dose rate dependencies were not observed when the intensity pattern was delivered using a conventional delivery technique. The observed distortion of the semi-Gaussian profile and dose variations among the segments of the four square intensity pattern are explained by considering the sampling rate and the communication time lag between the control systems. Finally, clinical significance is discussed

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
29
Journal Issue
3
Journal Page Range
p. 412-423
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35004190
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BEAM OPTICS; COLLIMATORS; DOSIMETRY; OPTIMIZATION; PHOTON BEAMS; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
BEAMS; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Notes
(c) 2002 American Association of Physicists in Medicine.