Published November 1, 1977 | Version v1
Journal article

A comparison of two methods of pulse-shape discrimination for alpha-gamma separation with trans-stilbene

  • 1. Ben-Gurion Univ. of the Negev, Beersheba (Israel). Dept. of Nuclear Engineering

Description

A method for measurement of low level alpha particles in high level gamma background is investigated. Because of its pulse-shape-discrimination properties and being a solid scintillator, trans-stilbene seems to be the proper scintillator, for this purpose. The investigation was done by measuring the effect of different gamma background level (from very low to very high) on constant alpha count rate. Two different pulse-shape-discrimination systems were used and compared. The Ortec system measures the pulse fall time and supplies a corresponding pulse height and the Elscint system checks whether the pulse is what is expected to be the gamma pulse, or is a longer pulse. Both systems yielded good results and were found to be adequate for alpha-gamma separation with trans-stilbene. (Auth.)

Additional details

Identifiers

Publishing Information

Journal Title
Nuclear Instruments and Methods
Journal Volume
146
Journal Issue
3
Series
Nucl. Instrum. Methods.
Journal Page Range
531-533
ISSN
0029-554X

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent