Utilization of glass detectors for high dose measurement

Description

Two kinds of glass were used for high dose measurement in 110 kCi Cobalt-60 irradiation facility. The correlation between absorbed doses from 0.2 to 80 kGy and optical densities of the dosimeters was studied. The optical density fading of the detectors at room temperature was measured over 60 days. The results show that glass detectors can be utilized as routine dosimeters with the main advantage of very low cost, simple measurement technique and wide dose range. (author). 8 figs, 6 refs, 2 tabs

Availability note (English)

Available from Information Centre (VINATOM), Hanoi (VN).
Part of:
High dose dosimetry in cobalt-60 gamma rays using glass detectors

Additional details

Additional titles

Original title (Vietnamese)
Su dung detector thuy tinh de do lieu cao

Publishing Information

Imprint Pagination
13 p.
Report number
INST--15/RR

INIS

Country of Publication
Viet Nam
Country of Input or Organization
Viet Nam
INIS RN
25018909
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Numerical Data, Non-conventional Literature
Descriptors DEI
ABSORPTION SPECTRA; EXPERIMENTAL DATA; GAMMA DOSIMETRY; ISODOSE CURVES; OPACITY; OPTICAL DEPTH CURVE; RPL DOSEMETERS
Descriptors DEC
DATA; DIAGRAMS; DOSEMETERS; DOSIMETRY; INFORMATION; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; NUMERICAL DATA; OPTICAL PROPERTIES; PHYSICAL PROPERTIES; SPECTRA

Optional Information

Notes
Report of the VINATOM's research project.