Published 1992
| Version v1
Report
Utilization of glass detectors for high dose measurement
Description
Two kinds of glass were used for high dose measurement in 110 kCi Cobalt-60 irradiation facility. The correlation between absorbed doses from 0.2 to 80 kGy and optical densities of the dosimeters was studied. The optical density fading of the detectors at room temperature was measured over 60 days. The results show that glass detectors can be utilized as routine dosimeters with the main advantage of very low cost, simple measurement technique and wide dose range. (author). 8 figs, 6 refs, 2 tabs
Availability note (English)
Available from Information Centre (VINATOM), Hanoi (VN).Additional details
Additional titles
- Original title (Vietnamese)
- Su dung detector thuy tinh de do lieu cao
Publishing Information
- Imprint Pagination
- 13 p.
- Report number
- INST--15/RR
INIS
- Country of Publication
- Viet Nam
- Country of Input or Organization
- Viet Nam
- INIS RN
- 25018909
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Numerical Data, Non-conventional Literature
- Descriptors DEI
- ABSORPTION SPECTRA; EXPERIMENTAL DATA; GAMMA DOSIMETRY; ISODOSE CURVES; OPACITY; OPTICAL DEPTH CURVE; RPL DOSEMETERS
- Descriptors DEC
- DATA; DIAGRAMS; DOSEMETERS; DOSIMETRY; INFORMATION; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; NUMERICAL DATA; OPTICAL PROPERTIES; PHYSICAL PROPERTIES; SPECTRA
Optional Information
- Notes
- Report of the VINATOM's research project.