Published December 1997 | Version v1
Journal article

Investigation of passivated silicon detectors for direct measurements of actinides

  • 1. CEA Centre d'Etudes de Fontenay-aux-Roses, 92 (France). Inst. de Protection et de Surete Nucleaire
  • 2. Canberra Semi-conductor NV, Olen (Belgium)
  • 3. CEA Centre d'Etudes de Saclay, 91 -Gif-sur-Yvette (France). Direction des Sciences de la Matiere

Description

Although much progress was made by using high purity germanium detectors for direct measurement of 239Pu in man, the improvement of lung spectroscopy sensitivity remains a challenge. Previous work showed that passivated implanted planar silicon detectors could provide an alternative by combining high-energy resolution at room temperature and planar geometry with large detection arrays. This paper reports on the status of this research based on the characterisation of 3x3 cm2 and 6x3 cm2 strip silicon detectors. Furthermore, the possibility to apply such detectors towards other monitoring problems for accidental situations such as the direct measurement of small contaminated surfaces was investigated. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
226
Journal Issue
1-2
Journal Page Range
p. 205-210.
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
9 refs.