Published September 1, 2009 | Version v1
Journal article

Experimental comparison of different dead-time correction techniques in single-channel counting experiments

  • 1. Bureau International des Poids et Mesures, Pavillon de Breteuil, 92312 Sevres Cedex (France)

Description

For single-channel counting, the dead-time correction method using transmission formulae was compared experimentally with the live-time technique. Agreement is within 4x10-4 for count rates up to 104 s-1 when the formulae for dead times in series are used and the dead-time value related to the analogue pulse width is adjusted for the BIPM modules. At higher count rates, a bias of 2x10-3 is observed. Two different live-time modules were compared and it was demonstrated that the live-time correction is also sensitive to the analogue pulse width and to the duration or rise time of the clock pulses. When these effects are taken into account, the two modules differ by only 5x10-4.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2009.06.010

Additional details

Identifiers

DOI
10.1016/j.nima.2009.06.010;
PII
S0168-9002(09)01200-5;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
608
Journal Issue
1
Journal Page Range
p. 163-168
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41045067
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COMPARATIVE EVALUATIONS; CORRECTIONS; COUNTING RATES; DEAD TIME; PULSE RISE TIME; PULSES; TRANSMISSION; WIDTH
Descriptors DEC
DIMENSIONS; EVALUATION; TIMING PROPERTIES