Published September 1, 2009
| Version v1
Journal article
Experimental comparison of different dead-time correction techniques in single-channel counting experiments
Creators
- 1. Bureau International des Poids et Mesures, Pavillon de Breteuil, 92312 Sevres Cedex (France)
Description
For single-channel counting, the dead-time correction method using transmission formulae was compared experimentally with the live-time technique. Agreement is within 4x10-4 for count rates up to 104 s-1 when the formulae for dead times in series are used and the dead-time value related to the analogue pulse width is adjusted for the BIPM modules. At higher count rates, a bias of 2x10-3 is observed. Two different live-time modules were compared and it was demonstrated that the live-time correction is also sensitive to the analogue pulse width and to the duration or rise time of the clock pulses. When these effects are taken into account, the two modules differ by only 5x10-4.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2009.06.010Additional details
Identifiers
- DOI
- 10.1016/j.nima.2009.06.010;
- PII
- S0168-9002(09)01200-5;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 608
- Journal Issue
- 1
- Journal Page Range
- p. 163-168
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41045067
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; CORRECTIONS; COUNTING RATES; DEAD TIME; PULSE RISE TIME; PULSES; TRANSMISSION; WIDTH
- Descriptors DEC
- DIMENSIONS; EVALUATION; TIMING PROPERTIES
Optional Information
- Copyright
- Copyright (c) 2009 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.