Published July 1988 | Version v1
Journal article

Some sources of error in thermoluminescent dosimetry

  • 1. Technical Univ. of Prague (Czechoslovakia). Faculty of Nuclear Science and Physical Engineering

Description

A wide variety of effects occurs during the use and evaluation of TL dosemeters causing possible erroneous reading of the absorbed dose. Main errors may be divided into two separate groups - the former being caused by the properties of the dosemeter, and the latter by instabilities of the TL reader. The paper summarizes the sources of error and evaluates some of them for LiF and aluminophosphate glass dosemeters. Instabilities caused by TL reader are closely associated with the method and frequency of calibration. The use of dosemeters irradiated by a known dose is time-consuming and reference light sources are commonly used for the operative control. Parameters of various scintillators used for these sources are discussed and the influence of their temperature dependence on the response is given. (author)

Additional details

Publishing Information

Journal Title
Kernenergie
Journal Volume
31
Journal Issue
7
Series
Kernenergie.
Journal Page Range
322-325
ISSN
0023-0642
CODEN
KERNA

Conference

Title
19. international symposium on radiation protection physics.
Dates
23-27 Mar 1987.
Place
Weinboehla (German Democratic Republic).