Published February 2010 | Version v1
Book

Development of air equivalent gamma dose monitor

  • 1. Electronics Division, Bhabha Atomic Research Centre, Mumbai (India)
  • 2. Electronics and Instrumentation Group, Bhabha Atomic Research Centre, Mumbai (India)

Description

The paper describes design and development of air equivalent gamma absorbed dose monitor. The monitor has gamma sensitivity of 84 pA/R/h for 60Co source. The characterization of the monitor has been done to get energy dependence on gamma sensitivity and response to gamma radiation field from 1 R/hr to 5000 R/hr. The gamma sensitivity in the energy range of 0.06 to 1.25MeV relative to 137Cs nuclide was within 2.5%. The linearity of the monitor response as a function of gamma field from 10 R/h to 3.8 kR/h was within 6%. The monitor has been designed for its application in harsh environment. It has been successfully qualified to meet environmental requirements of shock. (author)

Part of:
Proceedings of DAE-BRNS national symposium on nuclear instrumentation - 2010

Additional details

Publishing Information

Publisher
Dept. of Atomic Energy
Imprint Place
Mumbai (India)
ISBN
81-88513-33-4
Imprint Title
Proceedings of DAE-BRNS national symposium on nuclear instrumentation - 2010
Imprint Pagination
679 p.
Journal Page Range
p. 380-383

Conference

Title
national symposium on nuclear instrumentation
Acronym
NSNI 2010
Dates
24-26 Feb 2010
Place
Mumbai (India)

Optional Information

Notes
2 refs., 7 figs., 2 tabs.