Mean glandular doses for women undergoing mammographic breast screening in Oyo state, Nigeria
- 1. Department of Physics, University of Ibadan, Ibadan (Nigeria)
- 2. Radiology Department, University College Hospital, PMB 5116, Ibadan (Nigeria)
- 3. National Institute of Radiation Protection and Research, University of Ibadan, Ibadan (Nigeria)
Description
In this work, thermoluminescent dosimeters have been used to measure entrance surface doses (ESDs) of patients undergoing mammographic screening at the University College Hospital, Ibadan, Oyo state. The mean glandular doses (MGDs) were also calculated using the measured ESDs. The results showed that the ESDs ranged from 0.26 mGy to 21.26 mGy for the mediolateral oblique (MLO) views and 0.08 mGy to 5.36 mGy for the craniocaudal (CC) views. The calculated MGD ranged from 0.07 mGy to 3.57 mGy for the MLO views and 0.02 mGy to 0.98 mGy for the CC views. The possible reasons for the large variations in the individual ESD values and MGD values for both views are discussed using patients' data, equipment specific data and the technical parameters used for the examinations. Comparison showed that the mean ESD values and MGD values reported in this work are below published values. The mean of the calculated MGD values is also found to be lower than the recommended guidance level of 3.0 mGy when using grid. 92.5% of the patients had MGD values that are less than 2.5 mGy, hence a national reference MGD value of 2.5 mGy is proposed for Nigeria. Implementation of a dose reduction program in mammographic screening is also suggested because of the observed large variations in patients MGD values.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radi.2009.05.002Additional details
Identifiers
- DOI
- 10.1016/j.radi.2009.05.002;
- PII
- S1078-8174(09)00034-0;
Publishing Information
- Journal Title
- Radiography (London 1995)
- Journal Volume
- 15
- Journal Issue
- 4
- Journal Page Range
- p. 327-332
- ISSN
- 1078-8174
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41026831
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; MAMMARY GLANDS; PATIENTS; RADIATION DOSES; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; GLANDS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY
Optional Information
- Copyright