Analysis of X-ray beam parameters used to implement reference radiations for calibrating dosimetric systems for diagnostic radiology
Creators
- Oliveira, P.M.C.1
- Squair, P.L.1
- Nogueira, M.S.1
- Silva, T.A. da1
- International Radiation Protection Association (IRPA), Fontenay-aux-Roses (France)
- Sociedad Argentina de Radioproteccion (SAR), Buenos Aires (Argentina)
- International Atomic Energy Agency (IAEA), Vienna (Austria)
- Pan American Health Organization (PAHO), Washington, DC (United States)
- World Health Organization (WHO), Geneva (Switzerland)
- 1. Development Centre of Nuclear Technology (CDTN), Belo Horizonte, MG (Brazil)
Description
Ionizing radiation metrology is the base to achieve reliable dose measurements in all areas; it is also part of the framework of a radiological protection program that requires the use of reliable instruments that comply with standard performance criteria worldwide accepted. The International Electrotechnical Commission (IEC) established the X-ray reference radiations that are recommended to be used for calibrating dosimetric systems for diagnostic dosimetry. In this work, X-ray beam qualities were implemented in the Dosimeter Calibration Laboratory of the Development Centre of Nuclear Technology; the characteristics of the beams were analyzed through the measurement of parameters like inherent tube filtration, beam uniformity; beam field size, energy spectra and peak voltage. The influence of using 99.425% and 99.999% purity filters was also investigated. The first half-value layers and the homogeneity coefficients were measured for the three RQR2, RQR6 and RQR10 IEC beam qualities. An ionization chamber was also calibrated against a standard chamber in all implemented reference radiations and the relevant sources of uncertainties were estimated. Beam parameters were adjusted to comply with IEC standard; the main source of uncertainty during the calibration process was due to the ionization chamber positioning set-up and the purity of the filters did not cause relevant influence on the beam energy spectra. (author)
Files
41006633.pdf
Files
(142.4 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:0ed04f8d72dfb42bcfcf0d312b10e111
|
142.4 kB | Preview Download |
Additional details
Publishing Information
- Publisher
- SAR
- Imprint Place
- Buenos Aires (Argentina)
- Imprint Pagination
- 8 p.
- Report number
- INIS-AR-C--826
Conference
- Title
- 12. International congress of the International Radiation Protection Association (IRPA): Strengthening radiation protection worldwide
- Acronym
- IRPA 12
- Dates
- 19-24 Oct 2008
- Place
- Buenos Aires (Argentina)
INIS
- Country of Publication
- Argentina
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41006633
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CALIBRATION; DOSIMETRY; FILTERS; IONIZATION CHAMBERS; PHOTON BEAMS; RADIATION PROTECTION; X RADIATION
- Descriptors DEC
- BEAMS; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIATIONS; RADIOLOGY
Optional Information
- Notes