Published February 1988 | Version v1
Journal article

Physical performance of the latest generation of commercial positron scanner

  • 1. MRC Cyclotron Unit, Hammersmith Hospital, London (UK)

Description

A study is made of some of the basic physical performance characteristics of one of the latest generation of commercial positron tomographs. It consists of eight rings of BGO detectors (512/ring) with removable inter-ring septa and retractable transmission ring sources for each ring. The (transaxial) detector width is 5.6 mm and the lower limit of the spatial resolution in the reconstructed image is about 5 mm. Investigations were carried out into count rate linearity, image count recovery with object size, the effect of attenuation correction on quantitation and scatter fraction

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
35
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
721-725
ISSN
0018-9499
CODEN
IETNA

Conference

Title
34. nuclear science symposium and 19. nuclear power systems symposium.
Dates
21-23 Oct 1987.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-871006--.