Published June 1988
| Version v1
Journal article
Pocket sensitivity calibration of multicollector mass spectrometers
Creators
- 1. International Atomic Energy Agency, Vienna (Austria). Safeguards Analytical Lab.
Description
A method is described for the calibration of multicollector mass spectrometers. The instruments are used for nuclear application and equipped with 9 fixed collector pockets. To obtain all required information two solutions have been prepared one for uranium containing U-233, U-235, U-236 and U-238 with an approximate composition of 2:1:1:1 and the other for plutonium containing Pu-239, Pu-240, Pu-242 and Pu-244 with an approximate composition of 3:1:3:2. The spectrum was shifted by one and two mass units on both sides of the normal position and the ion intensities collected. These data have been used for the calculation of the relative sensitivity of the pockets involved. (orig.)
Additional details
Publishing Information
- Journal Title
- Fresenius' Z. Anal. Chem.
- Journal Volume
- 331
- Journal Issue
- 2
- Series
- Fresenius' Z. Anal. Chem.
- Journal Page Range
- 209-213
- ISSN
- 0016-1152
- CODEN
- ZACFA
Conference
- Title
- Symposium on trace element analysis by mass spectrometry.
- Original Conference Title
- Symposium ueber Massenspektrometrische Verfahren der Elementspurenanalyse
- Dates
- 30 Sep - 2 Oct 1987.
- Place
- Regensburg (Germany, F.R.).
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 19087009
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CONFIGURATION; EFFICIENCY; FARADAY CUPS; ISOTOPE RATIO; MASS SPECTRA; MASS SPECTROMETERS; MATHEMATICAL MODELS; PLUTONIUM ISOTOPES; SAFEGUARDS; URANIUM ISOTOPES
- Descriptors DEC
- BEAM MONITORS; MEASURING INSTRUMENTS; MONITORS; SPECTRA; SPECTROMETERS