Ceramic BeO exoelectron dosimeters for tritium and radon monitoring

Description

An environmental monitoring device with BeO ceramic dosimeters can be used to measure 222Rn in dwellings. Radon diffuses into a porous hemispheric chamber and the radon daughters are electrostatically collected on aluminized Mylar foil covering the BeO dosimeter that records the alpha activity. A 10:1 signal-to-background ratio results from a radon exposure of only pCi-h/l. This high sensitivity makes accurate radon measurement possible within one day, even at near background levels of a few tenths pCi/l. The BeO exoelectron dosimeter is also uniquely suited for monitoring occupational exposure to insoluble tritium gas. At one-fifth the maximum permissible concentration, exposure for 8 hours gives a 10:1 signal-to-background exoelectron response to the low energy beta rays. Compensation for any exoelectron response caused by photon radiation can be made by reading the thermoluminescence. The tritium exposure produces negligible thermoluminescence. Progress in these and other applications is now totally dependent on achieving reliability and long-term stability of the exoelectron dosimeter

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.

Files

11552960.pdf

Files (178.3 kB)

Name Size Download all
md5:882e5134c901a8f225595ddbbb8522f5
178.3 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
11 p.
Report number
CONF-7910149--1

Conference

Title
6. international symposium on exoelectron emission and dosimetry.
Dates
8 - 12 Oct 1979.
Place
Arhenshoop, German Democratic Republic.