Published 1975
| Version v1
Report
Characteristics of 130x130 mm2 drift chambers
Creators
- 1. Joint Inst. for Nuclear Research, Dubna (USSR)
Description
Standard multichannel drift chamber system tests have been carried out in order to optimize the drift chamber parameters. When a mixture of argon (70%) and isobuthane (30%) is used the following operating voltages are selected: 1.76 kV in anodic voltage and 2.8 kV in cathodic voltage. The tests have demonstrated that the chamber efficiency is above 99% and does not depend on the drift length. The calculations made with a regard to the test geometry indicate that the coordinate resolutions of chambers are at least 100 μm
Additional details
Additional titles
- Original title (Russian)
- Характеристики дрейфовых камер размерами 130x130 мм
Publishing Information
- Imprint Title
- Proportional and drift chambers
- Imprint Pagination
- p. 31-33.
- Report number
- JINR-D--13-9164
Conference
- Title
- International meeting on proportional and drift chambers.
- Dates
- 17 - 20 Jun 1975.
- Place
- Dubna, USSR.
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 9383840
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DRIFT CHAMBERS; EFFICIENCY; ELECTRIC POTENTIAL; OPTIMIZATION; SPATIAL RESOLUTION; SPECIFICATIONS
- Descriptors DEC
- MEASURING INSTRUMENTS; MULTIWIRE PROPORTIONAL CHAMBER; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RESOLUTION
Optional Information
- Notes