Published October 1995
| Version v1
Journal article
Dosimetric analysis on narrow 6 MV X ray beams
Description
The absorbed dose distribution of the small beams used in stereotactic radiotherapy has been measured. In a stereotactic treatment the target position is always centred on the accelerator isocenter. Therefore, when the absolute dose of the reference field size is known, only three parameters are essential to get a complete dosimetric determination: the tissue-maximum ratio, the output factor and the off-axis ratio. In this paper are reported the 6 MV dosimetric data of the above mentioned parameters for a series of circular collimators whose diameter ranges from 10 mm to 33 mm. Each parameter was obtained varying the measurement conditions and using different dosimeters: diodes, films and ionization chambers
Additional details
Identifiers
- PII
- 0167814096806839;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 37
- Journal Issue
- 3
- Journal Page Range
- p. S64
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34044471
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COLLIMATORS; DOSIMETRY; HARD X RADIATION; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; RADIATIONS; RADIOLOGY; THERAPY; X RADIATION
Optional Information
- Copyright
- Copyright (c) 1995 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.