Published June 2007
| Version v1
Journal article
Absorbed doses on patients undergoing tomographic exams for pre-surgery planning of dental implants
Creators
- 1. Centro de Desenvolvimento da Tecnologia Nuclear, Comissao Nacional de Energia Nuclear, Caixa Postal 941, CEP 30123-970 Belo Horizonte, MG (Brazil)
Description
The thermoluminescent (TL) dosimetry was used to measure entrance skin absorbed doses at anatomical points close to critical organs of patients undergoing tomographic techniques as part of a pre-surgery planning for dental implants. The dosimetric procedure was applied in 19 patients, and absorbed doses could be measured with a combined uncertainty down to 14%. Results showed that patient doses may be increased by a factor of 20 in the helical computed tomography compared to panoramic and spiral conventional tomographic exams
Additional details
Identifiers
- DOI
- 10.1016/j.apradiso.2007.01.016;
- PII
- S0969-8043(07)00037-1;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 65
- Journal Issue
- 6
- Journal Page Range
- p. 708-711
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 39007483
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; CRITICAL ORGANS; DOSIMETRY; PATIENTS; RADIATION DOSES; SURGERY
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; ORGANS; TOMOGRAPHY