Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance

  • 1. Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN 55905 (United States)

Description

Photon counting detectors (PCD) can provide spectral information to enable iodine quantification through multi-energy imaging but performance is limited by current PCD technology. The purpose of this work is to evaluate iodine quantification in a phantom study using dual-source PCD-CT (DS-PCD-CT), and compare to single-source (SS)-PCD-CT and traditional DS energy integrating detector (EID)-based dual-energy CT.

A multi-energy CT phantom with iodine inserts (0 to 15 mg ml−1 concentration) was imaged on a research SS-PCD-CT scanner (CTDIvol  =  18 mGy). A DS-PCD-CT was emulated by acquiring two sequential scans (CTDIvol  =  9 mGy each) using tube potentials: 140 kVp/80 kVp, 140 kVp/100 kVp and 140 kVp/120 kVp. For each kVp, 1 or 2 energy bins were reconstructed to achieve either dual-energy or quadruple energy CT. In addition to these energy combinations, a Sn filter was used for the high tube potential (140 kVp) of each kVp pair. For comparison, the same phantom was also scanned on a commercially available DS-EID-CT with matched radiation dose (CTDIvol  =  18 mGy). Material decomposition was performed in image space using a standard least-squares based approach to generate iodine and water-specific images. The root-mean-square-error (RMSE) measured over each insert from the iodine image was used to determine iodine accuracy.

The iodine RMSE from SS-PCD (140 kVp with 2 energy bins) was 2.72 mg ml−1. The use of a DS configuration with 1 energy bin per kVp (140 kVp/80 kVp) resulted in a RMSE of 2.29 mg ml−1. Two energy bins per kVp further reduced iodine RMSE to 1.83 mg ml−1. The addition of a Sn filter to the latter quadruple energy mode reduced RMSE to 1.48 mg ml−1. RMSE for DS-PCD-CT (2 energy bins per kVp) decreased by 1.3% (Sn140 kVp/80 kVp) and 15% (Sn140 kVp/100 kVp) as compared to DS-EID-CT.

DS-PCD-CT with a Sn filter improved iodine quantification as compared to both SS-PCD-CT and DS-EID-CT. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ab1c34

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
64
Journal Issue
11
Journal Page Range
[10 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52003872
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; ERRORS; FILTERS; IMAGE PROCESSING; IODINE; LEAST SQUARE FIT; PHANTOMS; PHOTON COUNTING; RADIATION DOSES
Descriptors DEC
DIAGNOSTIC TECHNIQUES; DOSES; ELEMENTS; HALOGENS; MATHEMATICAL SOLUTIONS; MAXIMUM-LIKELIHOOD FIT; MOCKUP; NONMETALS; NUMERICAL SOLUTION; PROCESSING; STRUCTURAL MODELS; TOMOGRAPHY