Image quality control of mammography equipment -Mammography System MX-300- of the Teachers Hospital of UNSA and dose measurement in breasts with radiographic films

  • 1. Universidad Nacional de San Agustin, Escuela Profesional de Fisica, Arequipa (Peru)

Description

This work is part of medical imaging for the evaluation of quality. Will have an accredited breast phantom Rmi-156 that allows evaluating the image quality of mammography equipment and through a series of techniques and processes that will submit to mammography films we obtain characteristic curves, which allows to evaluate different parameters that will serve for our study. Images were acquired with different k Vp and m As of the equipment, also with different thicknesses of the breast phantom. Also we want to use the lowest possible dose for obtaining our images. In this paper we develop a simple protocol that aims to unify the conditions under which are acquired the images for later evaluation. By obtaining these characteristic curves demonstrate that the Kodak film is the most suitable for our study because it requires lower dose for obtaining our images. (Author)

Files

47032297.pdf

Files (533.1 kB)

Name Size Download all
md5:ac2d5a035f274a6cf5fc6c2cbb21d176
533.1 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Control de calidad de imagen del equipo de mamografia -Mammography System MX-300- del Hospital de Docentes de la UNSA y medicion de dosis en mamas con peliculas radiograficas

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico, D. F. (Mexico)
Imprint Pagination
11 p.
Report number
INIS-MX--2961

Conference

Title
15. International Symposium on Solid State Dosimetry
Original Conference Title
15. Conferencia Internacional sobre Dosimetria de Estado Solido
Dates
26-30 Sep 2015
Place
Leon, Guanajuato (Mexico)

INIS

Country of Publication
Mexico
Country of Input or Organization
Mexico
INIS RN
47032297
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; DIAGRAMS; EVALUATION; FILMS; HOSPITALS; IMAGES; MAMMARY GLANDS; PHANTOMS; QUALITY CONTROL; RADIATION DOSES; THICKNESS; X-RAY EQUIPMENT
Descriptors DEC
BODY; BUILDINGS; CONTROL; DIAGNOSTIC TECHNIQUES; DIMENSIONS; DOSES; EQUIPMENT; GLANDS; INFORMATION; MEDICAL ESTABLISHMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS