Published August 1987 | Version v1
Miscellaneous

Dosimetry of new 137caesium sources before their application to clinical practice

  • 1. Institut pro Dalsi Vzdelavani Lekaru a Farmaceutu, Prague (Czechoslovakia)
  • 2. Ustav pro Vyzkum, Vyrobu a Vyuziti Radioisotopu, Prague (Czechoslovakia)

Description

A comparison is made of 137Cs sources of CMJ and CMT types of Czechoslovak production with similar CDC-J and CDCS-J sources manufactured by Amersham of Great Britain. Next to external dimensions the design was checked of the sources using autoradiography and X-ray radiography. The activity of the sources was measured and their quality determined (impurity content). Two calculation models were used for the calculation of dose distribution in the model and the obtained results were compared with published data. It was found that the Czechoslovak 137Cs source for clinical use has satisfactory parameters and its quality is comparable with the British source. (E.S.). 6 figs., 7 refs

Part of:
Topics in clinical oncology. 14

Additional details

Additional titles

Original title (no language)
Aktuality z klinickej onkologie. 14.

Publishing Information

Imprint Title
Topics in clinical oncology. 14
Imprint Pagination
196 p.
Journal Page Range
p. 82-86.
Report number
INIS-mf--11351

Conference

Title
2. symposium of radiological physicists.
Original Conference Title
II. symposium radiologickych fyzikov
Dates
22-25 Sep 1986.
Place
Smolenice (Czechoslovakia).

Optional Information

Notes
Imprint:Aktuality z klinickej onkologie. 14.