Published March 2003 | Version v1
Journal article

Measurement of absorbed dose and proposed radiation exposure level

  • 1. Tokai Univ., Isehara, Kanagawa (Japan). Hospital

Description

Absorbed dose was measured in clinical X-ray examinations using thermoluminescence dosimeter (TLD). Moreover, we distributed the levels of radiation exposure into 3 classes. The presumed dose of the internal organs, e.g., uterus dose, was computed to depth doses with a surface dose. This information provides a prediction of the influence of radiation, and the examination can be performed with the informed consent of the patient. Moreover, we examined the distribution of the level of absorbed dose. We proposed two kinds of radiation exposure level, one to the fetus in a pregnant woman and a general level of radiation exposure that is not applied to pregnant women. The levels were as follows: 0.5 mGy and 100 mGy were considered the boundaries for fetal radiation exposure in a pregnant woman, and 200 mGy and 3 Gy were considered the boundaries for the general level of radiation exposure (excluding pregnant women). (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
59
Journal Issue
5
Journal Page Range
p. 640-649
ISSN
0369-4305