Published May 2004
| Version v1
Journal article
Optimization study of a soft x ray flat field spectrometer
Creators
- 1. Chinese Academy of Sciences, Beijing (China). Inst. of Physics, Laboratory of Optics of Physics
Description
Using ray tracing code and optical path function, optimization calculation is carried out on a grazing incidence soft x ray flat field spectrometer with toroidal mirror as prepositive optic system. The authors calculate the spectral and spatial resolution of the spectrometer system. The wave front aberration theory and ray tracing code indicate that the system has good spectral resolution (λ/Δλ) better than 1000 and spatial resolution better than 40 μm in the spectrum range of 4-45 nm
Additional details
Publishing Information
- Journal Title
- Acta Physica Sinica
- Journal Volume
- 53
- Journal Issue
- 5
- Journal Page Range
- p. 1433-1439
- ISSN
- 1000-3290
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 36026786
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DESIGN; ENERGY RESOLUTION; OPTICAL SYSTEMS; OPTIMIZATION; SOFT X RADIATION; SPATIAL RESOLUTION; X-RAY SPECTROMETERS
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATIONS; RESOLUTION; SPECTROMETERS; X RADIATION