Published October 2018
| Version v1
Journal article
Native cardiac T1 Mapping: Standardized inline analysis of long and short axis at three identical 1.5 Tesla MRI scanners
- 1. Department of Radiology, University Hospital Erlangen, Maximiliansplatz 3, Erlangen, 91054 (Germany)
Description
Highlights: • High reproducibility of native cardiac T1 mapping independently of the heart segment. • Limited comparability between corresponding short and long axis segments. • To get reliable results multiple planes and axis should be acquired. • Follow-up examinations should be measured in identical segments. - Abstract: BackgroundInterpretation of native T1 time values remains difficult due to limited hard- and software comparability and the lack of measurement recommendations for the number of slices and their orientation. To provide a standardized comparison of native T1 time values in short and long axis and to investigate the interscanner reproducibility.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ejrad.2018.09.009Additional details
Identifiers
- DOI
- 10.1016/j.ejrad.2018.09.009;
- PII
- S0720048X18303152;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 107
- Journal Page Range
- p. 203-208
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51002661
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; COMPUTER CODES; HEART; MAGNETIC RESONANCE; NMR IMAGING
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; EVALUATION; ORGANS; RESONANCE