Published October 1999 | Version v1
Journal article

Beam-profile control using an octupole magnet

  • 1. Nuclear Science Research Facility, ICR, Kyoto University, Uji, Kyoto (Japan)

Description

For medical irradiation, beams with a uniform distribution have been required. We have proposed a method to control the beam profile using an octupole magnet. We have installed an octupole magnet and confirmed its effectiveness at the Heavy Ion Medical Accelerator in Chiba (HIMAC) beam-transport line. A beam efficiency of 91% can be realized with a flatness of ±4% over the range of ±15 mm. The beam offset at the position of the octupole magnet is required to be within ±0.2 mm to realize this condition. This method is expected to have a beam efficiency higher than that of the ordinary method using wobbler magnets and a scatterer system. (author)

Additional details

Publishing Information

Journal Title
Japanese Journal of Applied Physics. Part 1, Regular Papers, Short Notes and Review Papers
Journal Volume
38
Journal Issue
10
Journal Page Range
p. 6145-6149
ISSN
0021-4922