Published January 12, 2015 | Version v1
Journal article

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

  • 1. Liverpool and Macarthur cancer therapy centres and Ingham Institute, Liverpool Hospital, New South Wales (Australia)

Description

3D Gamma index is one of the metrics which have been widely used for clinical routine patient specific quality assurance for IMRT, Tomotherapy and VMAT. The algorithms for calculating the 3D Gamma index using global and local methods implemented in two software tools: PTW- VeriSoft® as a part of OCTIVIUS 4D dosimeter systems and 3DVHTM from Sun Nuclear were assessed. The Gamma index calculated by the two systems was compared with manual calculated for one data set. The Gamma pass rate calculated by the two systems was compared using 3%/3mm, 2%/2mm, 3%/2mm and 2%/3mm for two additional data sets. The Gamma indexes calculated by the two systems were accurate, but Gamma pass rates calculated by the two software tools for same data set with the same dose threshold were different due to the different interpolation of raw dose data by the two systems and different implementation of Gamma index calculation and other modules in the two software tools. The mean difference was -1.3%±3.38 (1SD) with a maximum difference of 11.7%

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/573/1/012054

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
573
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
8. International Conference on 3D Radiation Dosimetry
Acronym
IC3DDose
Dates
4-7 Sep 2014
Place
Ystad (Sweden)