Measurement of cardiac ventricular volumes using multidetector row computed tomography: comparison of two- and three-dimensional methods
- 1. Universite Victor Segalen Bordeaux 2, Laboratoire d'Anatomie Medico-Chirurgicale Appliquee, Bordeaux (France)
- 2. Hopital Haut-Leveque, Unite d'Imagerie Thoracique et Cardiovasculaire, Pessac (France)
- 3. Hopital Haut-Leveque, Service de Medecine Nucleaire, Pessac (France)
- 4. Universite Victor Segalen Bordeaux 2, Laboratoire de Physiologie Cellulaire Respiratoire C.R.I. INSERM 9806, Bordeaux (France)
- 5. Universite Victor Segalen Bordeaux 2, Departement de Radiologie, Bordeaux (France)
Description
This study compared a three-dimensional volumetric threshold-based method to a two-dimensional Simpson's rule based short-axis multiplanar method for measuring right (RV) and left ventricular (LV) volumes, stroke volumes, and ejection fraction using electrocardiography-gated multidetector computed tomography (MDCT) data sets. End-diastolic volume (EDV) and end-systolic volume (ESV) of RV and LV were measured independently and blindly by two observers from contrast-enhanced MDCT images using commercial software in 18 patients. For RV and LV the three-dimensionally calculated EDV and ESV values were smaller than those provided by two-dimensional short axis (10%, 5%, 15% and 26% differences respectively). Agreement between the two methods was found for LV (EDV/ESV: r=0.974/0.910, ICC=0.905/0.890) but not for RV (r=0.882/0.930, ICC=0.663/0.544). Measurement errors were significant only for EDV of LV using the two-dimensional method. Similar reproducibility was found for LV measurements, but the three-dimensional method provided greater reproducibility for RV measurements than the two-dimensional. The threshold value supported three-dimensional method provides reproducible cardiac ventricular volume measurements, comparable to those obtained using the short-axis Simpson based method. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-006-0222-5Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 16
- Journal Issue
- 10
- Journal Page Range
- p. 2341-2349
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 37111135
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; ERRORS; ISCHEMIA; MEASURING METHODS; THREE-DIMENSIONAL CALCULATIONS; TWO-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- ANEMIAS; CARDIOVASCULAR DISEASES; DISEASES; EVALUATION; HEMIC DISEASES; SYMPTOMS; VASCULAR DISEASES