Multicriteria optimization enables less experienced planners to efficiently produce high quality treatment plans in head and neck cancer radiotherapy

  • 1. Department of Radiation Oncology, University of Groningen, University Medical Center Groningen, PO Box 30001, Groningen, 9700 RB the (Netherlands)
  • 2. Department of Medical Imaging and Radiation Therapy, Hanze University of Applied Sciences, Groningen (Netherlands)

Description

To demonstrate that novice dosimetry planners efficiently create clinically acceptable IMRT plans for head and neck cancer (HNC) patients using a commercially available multicriteria optimization (MCO) system. Twenty HNC patients were enrolled in this in-silico comparative planning study. Per patient, novice planners with less experience in dosimetry planning created an IMRT plan using an MCO system (RayStation). Furthermore, a conventionally planned clinical IMRT plan was available (Pinnacle3). All conventional IMRT and MCO-plans were blind-rated by two expert radiation-oncologists in HNC, using a 5-point scale (1–5 with 5 the highest score) assessment form comprising 10 questions. Additionally, plan quality was reported in terms of planning time, dosimetric and normal tissue complication probability (NTCP) comparisons. Inter-rater reliability was derived using the intra-class correlation coefficient (ICC). In total, the radiation-oncologists rated 800 items on plan quality. The overall plan score indicated no differences between both planning techniques (conventional IMRT: 3.8 ± 1.2 vs. MCO: 3.6 ± 1.1, p = 0.29). The inter-rater reliability of all ratings was 0.65 (95% CI: 0.57–0.71), indicating substantial agreement between the radiation-oncologists. In 93% of cases, the scoring difference of the conventional IMRT and MCO-plans was one point or less. Furthermore, MCO-plans led to slightly higher dose uniformity in the therapeutic planning target volume, to a lower integral body dose (13.9 ± 4.5 Gy vs. 12.9 ± 4.0 Gy, p < 0.001), and to reduced dose to the contra-lateral parotid gland (28.1 ± 11.8 Gy vs. 23.0 ± 11.2 Gy, p < 0.002). Consequently, NTCP estimates for xerostomia reduced by 8.4 ± 7.4% (p < 0.003). The hands-on time of the conventional IMRT planning was approximately 205 min. The time to create an MCO-plan was on average 43 ± 12 min. MCO planning enables novice treatment planners to create high quality IMRT plans for HNC patients. Plans were created with vastly reduced planning times, requiring less resources and a short learning curve

Availability note (English)

Available from http://dx.doi.org/10.1186/s13014-015-0385-9; Available from http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4397887

Additional details

Publishing Information

Journal Title
Radiation Oncology (Online)
Journal Volume
10
Journal Page Range
vp.
ISSN
1748-717X

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47069957
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DOSIMETRY; GY RANGE 01-10; GY RANGE 10-100; HEAD; NECK; NEOPLASMS; OPTIMIZATION; PATIENTS; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
ABSORBED DOSE RANGE; BODY; DISEASES; DOSES; GY RANGE; MEDICINE; NUCLEAR MEDICINE; RADIATION DOSE RANGES; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) Kierkels et al.
Notes
PMCID: PMC4397887; PMID: 25885444; PUBLISHER-ID: 385; OAI: oai:pubmedcentral.nih.gov:4397887; licensee BioMed Central. 2015