Cooling rate effects in thermoluminescence dosimetry grade lithium fluoride
Creators
- 1. National Radiological Protection Board, Glasgow (UK). Scottish Centre
Description
A systematic investigation of the effects of cooling rates in the range 10-1 to 2 x 1050c min-1 applied to TLD-700, LiF thermoluminescence dosemeters has shown that the 'transfer sensitivity' effect observed by Booth, Johnson and Attix (1972, Health Phys., Vol. 23, 137) is only of importance for cooling rates > 1030c min-1. Although it is concluded that for practical dosimetry purposes the effect may be ignored until this discrepancy is explained it is recommended that a low temperature pre-irradiation anneal should be used. (author)
Additional details
Additional titles
- Subtitle (English)
- Implications for practical dosimetry
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine & Biology
- Journal Volume
- 21
- Journal Issue
- 1
- Series
- Phys. Med. Biol.
- Journal Page Range
- 60-66
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 7241142
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANNEALING; COOLING; ERRORS; LITHIUM FLUORIDES; PERFORMANCE; PERFORMANCE TESTING; SENSITIVITY; TEMPERATURE DEPENDENCE; THERMOLUMINESCENCE; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY; TIME DEPENDENCE
- Descriptors DEC
- ALKALI METAL COMPOUNDS; DOSEMETERS; DOSIMETRY; FLUORIDES; FLUORINE COMPOUNDS; HALIDES; HALOGEN COMPOUNDS; HEAT TREATMENTS; LITHIUM COMPOUNDS; LUMINESCENCE; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; TESTING