Published May 1990
| Version v1
Journal article
CsI(Tl) crystal scintillation detectors for computer tomography
Description
Uniformity of light yield and afterglow of the SDN.109 scintillation detectors for neurodiagnostics tomographs of the SRT 1010 second generation and the SDN.08 detectors for the SRT 5500 tomographs is studied. It is shown that 88% of detectors have millisecond afterglow of <0.1% and maximum sensitivity spread of combined detectors on their basis constitutes 8-12%. The scintillator dynamic range equals 108. Coefficients of radiation application are calculated and it is shown that CsI(Tl)-scintillator is efficient in combination with photomultipliers as well as with photodiodes
Additional details
Additional titles
- Original title (Russian)
- Сцинтилляционные детекторы на основе CsI(Тл)-кристаллов для компьютерной томографии
Publishing Information
- Journal Title
- Pribory i Tekhnika Ehksperimenta
- Journal Issue
- no.3
- Journal Page Range
- p. 198-201.
- ISSN
- 0032-8162
- CODEN
- PRTEAJ
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- Russian Federation
- INIS RN
- 24019723
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CESIUM IODIDES; COMPUTERIZED TOMOGRAPHY; PERFORMANCE TESTING; PHOTOMULTIPLIERS; SENSITIVITY; SOLID SCINTILLATION DETECTORS; SPECIFICATIONS; THALLIUM ADDITIONS