Published January 2010 | Version v1
Journal article

Characterization of a fiberoptic radiotherapy dosimetry probe based on Mg2SiO4:Tb

  • 1. Consejo Nacional de Investigaciones Cientificas y Tecnicas (CONICET) (Argentina)
  • 2. Instituto de Fisica Arroyo Seco-UNICEN, Pinto 399, 7000 Tandil (Argentina)
  • 3. Institute of Nuclear Sciences, Vinca, P.O. Box 522, 11000 Belgrade (Serbia)

Description

In this work the feasibility of using Mg2SiO4:Tb as a fiberoptic radioluminescent (RL) dosimetry probe for real-time dosimetry has been investigated for the first time. In particular, the stability of the RL signal after repeated use, the spectrum of the RL emission and the dose-rate response curve of a Mg2SiO4:Tb-based fiberoptic probe have been determined. The probe has been also used to obtain a percentage dose depth curve in a water phantom and its performance has been compared to that of a standard ion chamber. Besides, its absolute RL yield has been compared to that of an RL probe based on the commercial Al2O3:C phosphor.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2009.11.040