Published August 20, 2008 | Version v1
Journal article

Operation and investigation of a tilted bottom cavity for pyrgeometer characterizations

Description

A new cavity for pyrgeometer characterizations has been built at the Infrared Radiometry Section of the World Radiation Center (WRC-IRS) at PMOD/WRC. The calculated effective emissivity of 0.99993±0.00033 was obtained from Monte Carlo simulations taking into account the geometry and the measured temperature distributions of the cavity. The cavity is operated in a temperature range of -30 deg. C to +30 deg. C and is initially flushed with nitrogen to reduce the relative humidity in the cavity. The estimated uncertainties of retrieved pyrgeometer parameters k1, k2, and k3 are ±0.024, ±0.0008, and ±0.03, respectively. The relative uncertainty of pyrgeometer sensitivity C is 0.8%. The comparison with the cavity used since 1995 at PMOD/WRC gave average differences of 0.005, 0.00026, and 0.08 for k1, k2, and k3, respectively. The pyrgeometer sensitivity retrieved with the new cavity is on average 1.0% higher than with the original cavity

Additional details

Identifiers

Publishing Information

Journal Title
Applied Optics
Journal Volume
47
Journal Issue
24
Journal Page Range
p. 4441-4447
ISSN
0003-6935
CODEN
APOPAI

Optional Information

Notes
(c) 2008 Optical Society of America