Published January 2017 | Version v1
Journal article

Energy resolution and time resolution of SiPM coupled LYSO crystal detector

  • 1. Department of Nuclear Engineering and Technology, Chengdu (China)

Description

This paper studies the energy resolution and time resolution of scintillation detector produced by SEN-LE Company with appropriate measuring method. The detector is composed by LYSO crystal (model MicroFC-30035-SMT), and the coupling size of SiPM optical-electrical converter is 3 × 3 × 10 mm. The measuring result of the detector is: the energy of 22Na radioactive source is 511 keV, the energy resolution (FWHM) of 511 keV photon is 16%, and the time resolution is 135 ps. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
37
Journal Issue
1
Journal Page Range
p. 1-3, 19
ISSN
0258-0934

Optional Information

Notes
7 figs., 9 refs.