Published November 2014 | Version v1
Journal article

The correction of puncture point error in CT-guided puncturing biopsy: study of the methods

  • 1. Department of Medical Imagingy No.117 Hospital of PLA, Hangzhou, Zhejiang Province (China)

Description

Objective: To discuss the method of correcting the puncture point error in performing CT-guided puncturing biopsy. Methods: CT-guided puncture biopsy was performed in 201 cases on a GE Prospeed F II CT scanner. In order to easily understand the function relationship between bed coordinates and scan section coordinates, the gantry slant angles of 0°(n = 30), 5°(n = 50), 10°(n = 80) and 20°(n = 41) were used separately. Skin-marker was stuck on the skin to indicate the puncture point. CT-guided puncture/biopsy was carried out in all patients. CT scanning was repeated to adjust the puncture point error, and the actual error was estimated by water phantom experiment method. Results: A function relationship existed between bed coordinates and scan section coordinates. When the gantry position had no angle, the function formula was: TWM = S190 + M. When the gantry had some angle, the function formula was: TWM = S190/Cosθ + M. The function formula between the error value X and the gantry angle was: TWM = ∣190(1/Cosθ-1)∣. The error values obtained from function method were completely in agreement with the values determined by repeated CT scanning. Conclusion: Function method can precisely calculate the bed coordinates of target layer and quickly adjust the error of CT-guided puncture point. If these function formulas can be built in the scanner's software and worked automatically to adjust the error when the gantry is in angled state, the operation time as well as the radiation dose will be reduced. (authors)

Additional details

Publishing Information

Journal Title
Journal of Interventional Radiology
Journal Volume
23
Journal Issue
11
Journal Page Range
p. 972-976
ISSN
1008-794X

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
52083876
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOPSY; COMPUTERIZED TOMOGRAPHY; COORDINATES; CORRECTIONS; ERRORS; LAYERS; PATIENTS; PHANTOMS; RADIATION DOSES; SKIN
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DOSES; MOCKUP; ORGANS; STRUCTURAL MODELS; TOMOGRAPHY

Optional Information

Notes
6 figs., 1 tab., 8 refs.; http://dx.doi.org/10.3969/j.issn.1008-794X.2014.11.009